15W Medical Grade Desktop Switching Power Adapters AMP1501 series
WWW.ASTRODYNE.COM ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
Universal 90-264VAC Input
2 Input Connection Options
Single Outputs from 5 to 36VDC
4000V Isolation
LPS Efciency Level V
Splash Proof Case
Model Number Output Watts Output Voltage Output Amps Total Regulation (See Notes)
*AMP1501-02(A or B) 13 5V 2.6A 5%
*AMP1501-03(A or B) 13 7V 1.85A 5%
*AMP1501-04(A or B) 15 9V 1.66A 5%
*AMP1501-05(A or B) 15 12V 1.25A 5%
*AMP1501-06(A or B) 15 15V 0.93A 5%
*AMP1501-07(A or B) 15 18V 0.83A 5%
*AMP1501-08(A or B) 15 24V 0.62A 3%
*AMP1501-09(A or B) 15 30V 0.5A 3%
*AMP1501-10(A or B) 15 36V 0.41A 3%
Notes:
*Denotes a choice of AC input connectors (A or B option). Please see Mechanical Dimensions. If not denoted, option “A” IEC-320-C14 input
connector is used.
On AMP1501-02, 03 and 04 a 16AWG/4FT output cable is necessary to achieve the Total Regulation specication listed. 18AWG/4FT is
required for all other models.
Low cost 3 Wire IEC320/C13 (Part Number ACCRD-C13) and 2 Wire IEC320/C7 (Part Number ACCRD-C7) Input Power Cords are available for
these products. Inquire to the Astrodyne Sales Depart.
Option “A”
Option “B”
4.11” x 1.65” x 1.22”
15W Medical Grade Desktop Switching Power Adapters AMP1501 series
WWW.ASTRODYNE.COM ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
INPUT SPECIFICATIONS
Input Voltage Range 90-264 VAC (Universal)
Frequency Range 47-63 Hz
Input Current Max. Low Line (115VAC) 0.33A
High Line (230VAC) 0.18A
No Load Power Consumption 0.3W @ 230VAC
Inrush Current, typ (Cold Start) Low Line (115VAC) 16A max.
High Line (230VAC) 30A max.
Leakage Current 0.3mA MAX.
@ 240 VAC, FL
OUTPUT SPECIFICATIONS
Voltage and Current See Selection Charts
Output Power Range 0 to FL Rating
Load Regulation (230Vin) ± 5%, max.
Line Regulation ± 1%, max.
Transient Response 100Vin, FL to 1/2L 4mS
Temperature Coefcient ±0.04%/°C, typ
Ripple/Noise FL, 90Vin 1% Pk-Pk
Over Voltage Protection 112-132%
Over Current Protection 110-150%
Short Circuit Protection Continuous, self-recovering *
Thermal Shutdown -20 to 125°C
Start Up / Hold Up Time 100Vin, FL 1S max. / 10 mS min.
GENERAL SPECIFICATIONS
Input-Out Isolation 5656VDC (4000VAC)
Isolation Resistance 50 MΩ min.
Efciency (FL, 230Vin) 75%, typ
Safety (c)UL60601-1, EN60601-1
PHYSICAL SPECIFICATIONS
Size 4.11” x 1.65” x 1.22”
Construction Desktop Style
Weight 6.0oz (170g)
ENVIRONMENTAL SPECIFICATIONS
Oper. Temperature 0 to +50°C (FL) *
50% Load @ 70°C
(See Derate Curve)
Storage Temperature -40 to +85°C *
Relative Humidity 5% to 95%, non-cond *
EMC 110Vin: FCC Part 15 Class B
EN55022B
MTBF 100,000 Hrs
Efciency Level V
DERATE CURVE
0 0 0 0 0 0 0 0
0
0
0
0
0 0
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
All specications are typical at nominal input, full load, and 25°C unless other-
wise noted Astrodyne products are not authorized or warranteed for use
as critical components in life support systems, equipment used
in hazardous environments, nuclear controls systems, or other
mission-critical applications.
WWW.ASTRODYNE.COM ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
15W Medical Grade Desktop Switching Power Adapters AMP1501 series
MECHANICAL DIMENSIONS - AMP1501-XXA
DC OUTPUT CONNECTIONS
Coaxial Plug
+Vout-Vout
5.5mm outer diameter
2.1mm inner diameter, female
11mm length, center “+”
Output Cord Length 48” typ.
WWW.ASTRODYNE.COM ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
15W Medical Grade Desktop Switching Power Adapters AMP1501 series
MECHANICAL DIMENSIONS - AMP1501-XXB
DC OUTPUT CONNECTIONS
Coaxial Plug
+Vout-Vout
5.5mm outer diameter
2.1mm inner diameter, female
11mm length, center “+”
Output Cord Length 48” typ.